Correcting the effect of hematocrit in whole blood coagulation analysis on paper-based lateral flow device
Document Type
Article
Publication Title
Analytical Methods
Abstract
A simple method to correct the effect of hematocrit (Hct) on a paper-based lateral flow assay device that is designed to assess whole blood coagulation is reported. By quantifying the effect of Hct on red blood cell (RBC) transport, a calibration procedure is obtained that compensates for the Hct effect, while maintaining the dependence of blood coagulation on RBC transport.
First Page
2869
Last Page
2874
DOI
10.1039/c8ay00192h
Publication Date
6-28-2018
Recommended Citation
Li, Hua; Han, Daewoo; Pauletti, Giovanni M.; Hegener, Michael A.; and Steckl, Andrew J., "Correcting the effect of hematocrit in whole blood coagulation analysis on paper-based lateral flow device" (2018). Pharmaceutical and Administrative Sciences Faculty Publications. 142.
https://doi.org/10.1039/c8ay00192h
https://collections.uhsp.edu/pharm-admin-sciences_pubs/142